PLT Unit – pipette leak testing unit

The most frequent cause of inaccuracy in air displacement pipettes is leakage. This arises from damage either to seals, pistons, or tip cones. Often not detectable by the naked eye, leaks lead to significant volume errors. The Brand pipette leak tester (PLT unit) for air displacement pipettes finds even the smallest leak within seconds. To monitor measuring instruments according to ISO 8655, air-displacement pipettes must be checked at regular intervals and the results must be compared with the error limits contained in ISO 8655-2. However, a calibration certificate only reflects the performance of the instrument at the time of testing. The time between calibrations is crucial, since leaks can occur at any time, even immediately after testing. Well over 80% of pipettes sent in for repair have leaks and are outside their volume tolerances, even if they don’t drip.

Features

  • Test limit values for all commercially available single- and multichannel pipettes in the volume range 1 ?l to 10 ml are pre-programmed.
  • Test with and without tips to assist in leak source diagnosis
  • Test results in seconds
  • Patent pending